Brand Name :STATEMED
Model Number :STT15-MPM44
Certification :CE
Place of Origin :CHINA
MOQ :1 SET
Price :competitive
Payment Terms :L/C, T/T, Western Union
Supply Ability :100 sets per month
Delivery Time :10 working days
Packaging Details :carton, wooden case
Display mode :LCD display
Product Name :Large screen patient monitor
Battery type :Lithium battery 11.1V/4.4Ah
Recharge time :≥4 hour
Use time :3 hour
Environment temperature :5℃~+40℃
more
Contact Now

Add to Cart

Find Similar Videos
View Product Description

Doctor Diagnose Portable Multi Parameter Patient Monitor CE Approved

Battery

Battery typeLithium battery 11.1V/4.4Ah

Recharge time
(90% charge time)

≥4 hour
Use time3hours



Work Environment

Environment temperature

Working 5℃~+40℃
Transportation and storage -40℃~+55℃

Relative humidity

Working ≤85%
Transportation and storag ≤95%

Atmospheric pressure

Working 700hpa~1060hpa
Transportation and storag 500hpa~1060hpa



ECG

LeadStandard Three leadsR,L,F, lead mode :Ⅰ,Ⅱ,Ⅲ
Standard Five leads

R,L,F,N,C,
Lead mode :Ⅰ,Ⅱ,Ⅲ,aVR,aVL,aVF,V;

Sensitivity2.5mm/mV,5mm/mV,10mm/mV,20mm/mV,40mm/mV,erro≤±3%
Scan speed12.5,25,50mm/s three level,erro ≤±10%
Adjustable screen aspect ratio0.05~3.2
Lead off condition detectedU=2V I=0.04uA
Active noise suppressionIn patients who have not applied current
Tall T-wave suppression1.4mV
Inner voice<30μVP-P
Input impedance≥2.5MΩ
Input loop current≤0.1μA
Endure electrode polarizing voltage±300mV
ECG signal input range±5mV
Measurement modeDiagnosis mode,monitoring mode,operation mode
Frequency property

diagnosis mode 0.05~130HZ
monitoring mode 0.5~40HZ
operating 1~20HZ

CMRR

Diagnosis mode ≥80dB
Monitoring mode ≥90dB
Operating mode ≥100dB

HR measurement range

Adult 30~250bpm
Neonate /pediatrics 30~300bpm

HR accuracy±2%
HR sensitivity>200μV
Heart rate calculationTake 4 to 12 R-wave cycle averaged
Refresh time heart rate1S
Heart rate response time

80-120 BPM:range 4-7s
Average 4-7s
80-40 BPM: range 4-7s
Average 4-7s

Alarm range0~301bpm,upper and lower limit can be adjusted continuously
Alarm indicatorAudial and visual alarm
Heart rate too fast startup time of the alarm

Range :4-6s
Average :4-6S

ProtectionIsolated immunity voltage 4000V,50/60HZImmunity of electronic surgical Anti-defibrillation
Protection and recovery for defibrillation effectIn the monitor and the surgical model ,ECG waveforms recovery to baseline in 5S.
Electrode polarization recovery time after the fibrillationIn the monitor and the surgical mode ,ECG waveforms recovered to baseline in 10S.
Surgical interference of electrical recovery timeIn the monitor and the surgical mode ,ECG waveforms recovered to baseline in 10S.
No overshoot pacemaker pulse suppression:
Extent±2mV~±700mV,Pulse width 0.1ms~2msQRS waveDisplay
a)Single pacing pulse6060±2bpm
3030±2bpm
0--
b)With normal QRS-T pacing the pacing pulse6060±2bpm
3030±2bpm
0--
c)QRS has an invalid pacing mode pacing pulse. In all cases, overshoot (a0) pacing rate should be less than 5% overshoot and settling time must be less than 5μs. Rise and fall time of the pulse width should be 10% but not more than 100μs, and the pacing pulse starting time should be the starting point in the QRS complex prior to or shorter than 40ms at 40ms.6060±2bpm
3030±2bpm
0--
Beware of the room pacing pulses have the same ventricular pacing pulse amplitude and duration, and ventricular pacing pulses than 150ms to 250ms in advance6060±2bpm
3030±2bpm
0--
Typical minimum input slew rate50V/s RTI