High quality hospital devices portable light hospital examining lamps
1. Product Show:
Characteristic:
It can be used with wall clip, table clip and floor prop. It has features of simple operation, easy to install, low-voltage and safety . It can be used in the clinics of general hospital , wards, departments and community hospitals and army field hospitals and so on.
Product component:
Mainly by the lamp holder, light poles, power boxes and other components.
Performance Summarize:
The examination lamp is designed with wide input voltage, low voltage output power supply, has a few characteristics of stable performance, safe and reliable, widely used.
The design and production of this light source strictly executes the first part of the safety requirements of medical electrical equipment GB4793.1-2007: general requirements which is guaranteed in safety.
Main technical parameters:
1. Voltage:AC90~240V 50-60Hz
2. Input power:≤30VA
3. Bulb power:LED 9W
4. Color temperature:5500±500K
5. Life of bulb:50000h
6. Illumination:30000lx
7. N.W.:1.44KG
8. Package dimension(with clip):600*120*260mm (with floor prop):845*215*155mm
9. Packaging material: Carton
Security tips and symbol instructions
1.When the examination lamp is in use, you should always abide by the basic safety precautions in order to avoid fire, electric shock or the risk of injury,.
2.Please read the instructions carefully, and operating according to the specifications and requirements.
3.The replacement of fuse protector should be the same with the mark.
4.When no use, please unplug the power cord.
Working Environment
Environmental temperature: 5℃~40℃
Relative humidity : ≤85%
The atmosphere: 86.0~106.0kPa
Transportation and storage conditions
A. Environmental temperature: -40℃~55℃
B. Relative humidity : ≤90%
C.The atmosphere: 50Kpa~106Kpa
Failure Judgment and Eliminating Methods
(It should be operated by the professionals)
Fault phenomena | Fault cause | Solution |
The light is not bright | 1.Power lines have no electricity | Connect to the circuit where there is electricity supply |
2.The power cord is not connected well | Reconnect the power cord | |
3.Brightness adjusting button is at the minimum position | Clockwise potentiometer rotation to the end | |
4.Light spot adjusting button is not open | Open the Light spot adjusting button | |
5.Fuse damage | Change the fuse | |
6.Power Board damage | Return to the factory for maintenance |