Medical Liquid Nutrition Pump Enteral Nasogastric Feeding Pump
Main Technique Parameters:
a) Infusion rate:0ml~999ml,≤±10%
b) Total volume:0.1ml~999ml,≤±10%
c) flow rate:1ml/h~300ml/h;
d) Infusion accuracy:At 1ml/h~50ml/h,range:1ml/h,≤±10%;
At 50ml/h~300ml/h,range 5ml/h,≤±10%
e) Power consumption:13VA
f) Size: 193mm×130mm×105mm
g) N.W:1.6kg
h) Feeding tube:Special feeding tube
Working Conditions:
a) Ambient temperature:5℃~40℃;
b) Relative humidity:30%~80%;
c) Barometric:860~1060hPa;
d) Power supply:AC220V,50Hz
e) Battery:Ni-H2 battery group 9.6V,1600mAh
Alarm function:Sound and light alarm as following situations
a) Without any operating when switched on
b) when reach reques volume
c)low battery
d)Feeding tube is in empty,blocked and folding
e) When working loss of AC supply suddenly
Problem | Possible Causes | What To Do |
Alarm Continue | 1)Liquid above line in “venture” catheter” 2)Bodies block monitoring port | 1)Lowering water 2)removal the bodies(Maybe water droplets in monitoring, can swing tube) |
Interrupt alarm | Pump in forward or backward place | Put in flat place |
After starting pump, not rotate and interrupt light on | Pump wheel is stuck by nutritious | Unscrew pump wheel and clean it. |
When pump intermittent stall, “venturi catheter” is still dripping | The tube is too long or joint is slide out from position | 1)Choose special feed tube.2)recovery joint in original position |
Flow rate is inaccurate | Didn't use special feeding tube | Use special feeding tube |