Add to Cart
128 E Diagnostic Baby Color Doppler Ultrasound Scanner B / M Display Mode Customized
Performance requirements in grayscale imaging mode
Color ultrasound in the gray scale imaging mode of the performance of the probe:
Performance | Probe type and nominal frequency | ||
2.0≤f<5.0 | 5.0≤f<8.0 | 5.0≤f<10.0 | |
a) Robe type, model | Convex array (model TC60A) | Cavity (model TC10A) | Line array (model TL40A) |
b) Robe nominal frequency (MHz) | 3.5 | 5.5 | 7.5 |
c)Robe depth(mm) | ≥160 | ≥80 | ≥50 |
d) Lateral resolution (mm) | ≤3(depth≤80) ≤4(80<depth≤130) | ≤2(depth≤40) | ≤2(depth≤40) |
e) Axial resolution (mm) | ≤2(depth≤80) ≤3(80<depth≤130) | ≤1(depth≤40) | ≤1(depth≤50) |
f) Blind zone (mm) | ≤5 | ≤5 | ≤3 |
g) Horizontal geometric position accuracy (%) | ≤15 | ≤20 | ≤10 |
h) Longitudinal geometric position accuracy (%) | ≤10 | ≤10 | ≤5 |
i) Slice thickness (mm) | ≤7 | ≤5 | ≤5 |
j) Perimeter and area measurement deviations (%) | ≤±20 | ≤±20 | ≤±20 |
k) M mode time display error (%) | ≤±10 | ≤±10 | ≤±10 |
Transport: